Radiant energy – Inspection of solids or liquids by charged particles – Electron probe type
Reexamination Certificate
2000-07-07
2002-07-16
Nguyen, Kiet T. (Department: 2881)
Radiant energy
Inspection of solids or liquids by charged particles
Electron probe type
C250S307000, C250S252100, C257S048000
Reexamination Certificate
active
06420702
ABSTRACT:
FIELD OF THE INVENTION
This invention deals with metrology standards, and in particular with metrology standards as applied to Scanning Electron Microscopy used for measurement of Critical Dimensions (CD's) in integrated circuit manufacturing technology.
BACKGROUND OF THE INVENTION
As integrated circuits become smaller and faster, the critical dimensions (CD's) of the devices and interconnections also must decrease. As these CD's get closer to the resolution limits of optical lithography and microscopy measurement techniques, great care must be taken to eliminate all possible sources of measurement error in order to obtain accurate and reproducible critical dimensions. One nearly universally used measurement technique is Scanning Electron Microscopy (SEM), which utilizes highly focused energetic beams of electrons impinging on the sample and measures the yield of secondary emitted electrons. SEM is the most widely used tool for VLSI measurement and morphology analysis, due to its high resolution and relative ease of use.
FIG. 1
depicts an SEM system, showing the electron source and the acceleration, focusing, and detection electronics.
FIG. 2
shows a typical electron emission energy spectrum resulting from the incident electron beam of an SEM. The highest energy peak
18
results from the backscattered electrons, which have energies close to that of the incident beam, and which have undergone only elastic collisions with the target atoms. Peaks
20
seen at intermediate energies are the Auger electrons emitted due to relaxation of electrons between atomic energy levels. The lowest energy emitted electrons
22
, produced by inelastic collisions between the primary beam and the inner shell electrons of the sample, are known as the secondary electrons and are generally the most useful for morphology studies in VLSI. This is due in part to the extremely short escape depth (less than about 50 Angstroms) of secondary electrons, which yields high surface sensitivity. In addition, since the incident electron beam undergoes beam broadening due to multiple collisions as it penetrates into the sample, the backscattered electrons originating from deeper into the sample reflect this broadening with degraded point-to-point resolution. The lower energy secondary electrons which escape the sample originate from the surface region above the penetration depth where beam broadening becomes influential, and therefore yield higher point-to-point resolution than evidenced by backscattered electrons.
The detected electron current, typically chosen to be the secondary electron current as described above, is used to intensity modulate the z-axis of a CRT. An image of the sample surface is produced by synchronously raster scanning the CRT screen and the electron beam of the SEM.
The contrast of the image depends on variations in the electron flux arriving at the detector, and is related to the yield of emitted electrons per incident electron. The yield is dependent on both the work function of the material and the surface curvature. These factors allow the SEM to distinguish between materials such as metal, oxide, and silicon, and also to distinguish surfaces which differ in slope. Thus, Critical Dimensions (CD's) of patterned and/or etched lines and gaps can be measured.
Two factors affecting the accuracy of SEM measurements are resolution and charging effects. The resolution of the SEM depends on the type of sample under inspection and on the incident beam diameter or “spot size”. The high voltages of the electron beam required to produce small scanning spot sizes were historically one source of charging when examining insulating surfaces. When incident beam energies exceeded the secondary electron crossover point, i.e., when the incident beam penetration depth was high enough that the number of emitted secondary electrons was less than the number of incident electrons, the surface in the region of the scanning beam would acquire excess negative charge, which would be retained in the case of exposed insulating surfaces. This would cause the incident beam trajectory to be disturbed, and would therefore degrade the image. Grounding schemes such as coating the surface with a conducting material, e.g., gold, and attaching a ground wire to the coating were used to reduce charging effects. These methods prohibited further processing following inspection of wafers. More recent SEM machines have eliminated high energy accelerating voltages, thus decreasing the corresponding charging effects. However, these low voltage SEM's have poorer resolution (approximately 40-50 A resolution for 600 V SEM) than the corresponding high-voltage SEM's (approximately 5-10 A for 20-50 KV SEM). Charging of insulating surfaces can result from factors other than the SEM electron beam, such as processing-induced charging.
Calibration standards are necessary when using SEM's, in order to assure that magnification factors are accurate and to thereby assure accurate sizes as measured by SEM photographs. Conductor-coated (generally gold-coated) standard samples with features measured to high accuracy are still used to calibrate, i.e., to verify CD's and spacings as measured in low-voltage SEM's. Although this method is one way of producing measurement standards, it has the disadvantage of degrading image contrast, since the underlying layers are partially obscured by the conducting coating. Additionally, gold-coated standards cannot be used in fabrication facilities since the presence of gold incurs the risk of drastically lowering the carrier lifetime in silicon.
If measurement standards having exposed insulating layers are used without a conducting coating in order to achieve maximum contrast, variations in charging will induce variations in measurement, and therefore the calibration standard will drift as a function of time.
In order to achieve a calibration standard which will be effective for measuring linewidths less than 0.1 microns, for polysilicon by way of example, it is beneficial that the standard have a grainy structure at the top surface of the features (which provides focus improvement for the e-beam) while not being grainy on the feature sidewalls (which would result in measurement variations along the line), that the standard has a vertical and smooth sidewall profile, that it can be patterned as finely as the linewidths being measured, that it is not susceptible to charging effects, and that it yields good image contrast. Currently available SEM standards do not fulfill these requirements, and in particular are generally subject to significant charging effects.
SUMMARY OF THE INVENTION
It is therefore an object of this invention to provide an improved SEM measurement standard which is not subject to surface charging.
It is a further object of this invention to provide an improved SEM measurement standard which improves image contrast and e-beam focus.
It is a further object of this invention to provide an improved SEM measurement standard which can be patterned to linewidths of 0.1 microns or less, with vertical and smooth sidewall profile.
These objects are met by creating an SEM measurement standard having two different conducting layers comprised of materials used in semiconductor fabrication.
REFERENCES:
patent: 5414265 (1995-05-01), Sartore
patent: 5434409 (1995-07-01), Tsubusaki
patent: 5637186 (1997-06-01), Liu et al.
patent: 5977542 (1999-11-01), Singh et al.
patent: 6048743 (2000-04-01), Yang et al.
patent: 6057171 (2000-05-01), Chou et al.
patent: 6137175 (2000-10-01), Tabara
patent: 6215129 (2001-04-01), Harvey et al.
Singh Bhanwar
Templeton Michael K.
Tripsas Nicholas H.
Fisher Gerald
Nguyen Kiet T.
Wenocur Deborah
LandOfFree
Non-charging critical dimension SEM metrology standard does not yet have a rating. At this time, there are no reviews or comments for this patent.
If you have personal experience with Non-charging critical dimension SEM metrology standard, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Non-charging critical dimension SEM metrology standard will most certainly appreciate the feedback.
Profile ID: LFUS-PAI-O-2848891