Broadband spectroscopic rotating compensator ellipsometer

Optics: measuring and testing – By polarized light examination – Of surface reflection

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

C356S365000

Reexamination Certificate

active

06320657

ABSTRACT:

FIELD OF THE INVENTION
The present invention relates to ellipsometers, and more particularly to rotating compensator ellipsometers.
BACKGROUND OF THE INVENTION
Optical ellipsometry has long been recognized as providing accurate characterizations of semiconductors and other materials, their surface conditions, layer compositions and thicknesses, and for characterizing overlying oxide layers. This non-destructive technique is particularly needed to evaluate thickness, crystallinity, composition and index of refraction characteristics of thin films deposited on semiconductor or metal substrates to ensure high yields during fabrication.
An ellipsometer probes a sample with a light beam having a known polarization state. The light beam is reflected at a non-normal incidence from the surface of the sample. The polarization state of the beam is modified upon reflection in a way that depends upon the structure of the sample. By accurately measuring the polarization state of the reflected beam and comparing it to the original polarization state, various properties of the sample can be ascertained.
In spectroscopic ellipsometry, the probing wavelength is changed and the ellipsometric measurement is repeated at each new wavelength.
Spectroscopic ellipsometry is ideal for multi-material samples formed in stacked layers. The different depth penetrations and spectral responses that vary depending upon the material and wavelength of light provide additional information about a sample that is not available using single wavelength ellipsometers.
One possible ellipsometric configuration uses the following components: 1) a light source, 2) a linear polrize (“polarizer”), 3) the sample (to be analyzed), 4) a second linear polarizer (“analyzer”), and 5) a detector. The polarizer produces totally polarized light, and the analyzer assesses the polarization state of the beam after reflection from the sample. Either the polarizer or the analyzer is rotated so that the detector signal can be used to accurately measure the linear polarization component of the reflected beam. Then, the circularly polarized component is inferred by assuming that the beam is totally polarized, and what isn't linearly polarized must be circularly polarized.
An advantage of this type of ellipsometer is that the polarizer and analyzer can function over a wide range of wavelengths, thus providing a broad perspective of the sample as compared to measurements made at a single wavelength. A common method of spectroscopic ellipsometry is to use a broadband light source, such as a Xe arc lamp, with data acquisition occurring in a serial or parallel mode. In the serial mode, a monochrometer is placed before the detector to filter out all wavelengths except for the desired probe wavelength. Multi-wavelength measurements are performed serially, one wavelength at a time, by properly adjusting the monochrometer for each measurement. In parallel operation, a polychromatic light beam having a broad range of wavelengths is directed to the sample. The reflected beam is diffracted to a photodetector array or an optical multichannel analyzer by a diffraction grating or a prism. The multi-wavelength measurements are performed in parallel, with all wavelengths of interest being measured at the same time.
Such an ellipsometer, commonly called a rotating-polarizer or rotating-analyzer ellipsometer, is termed “an incomplete” polarimeter, because it is insensitive to the handedness of the circularly polarized component and exhibits poor performance when the light being analyzed is either nearly completely linearly polarized or possesses a depolarized component.
The latter limitations of the rotating-polarizer and rotating-analyzer ellipsometers can be overcome by including a rotating compensator placed between the polarizer and the analyzer (both of which are fixed and not rotating) that rotates about the propagating axis of the beam. The compensator can be placed either between the sample and the polarizer, or between the sample and the analyzer. Such a configuration is commonly called a rotating compensator ellipsometer. The compensator is an optical component that delays the light polarized parallel to its slow axis relative to light polarized parallel to its fast axis by an amount proportional to the refractive index difference along the two directions and the thickness of the plate, and inversely proportional to the wavelength of the light.
It is known in the art that the intensity I of a beam transmitted through an ideal compensator-analyzer combination is expressed as:
I

[
&LeftBracketingBar;
E
x
&RightBracketingBar;
2
+
&LeftBracketingBar;
E
y
&RightBracketingBar;
2
]
=
I
o
[


&LeftBracketingBar;
E
x
&RightBracketingBar;
2

(
cos
2

(
δ
/
2
)
+
(
1
/
2
)



sin
2

(
δ
/
2
)
)
+
&LeftBracketingBar;
E
y
&RightBracketingBar;
2

(
1
/
2
)



sin
2

(
δ
/
2
)
+
(
&LeftBracketingBar;
E
x
&RightBracketingBar;
2
-
&LeftBracketingBar;
E
y
&RightBracketingBar;
2
)

(
1
/
2
)

sin
2

(
δ
/
2
)



cos

(
4

C
)
+
Re



(
E
x

E
y
*
)

sin
2

(
δ
/
2
)



sin



(
4

C
)
-
Im

(
E
x

E
y
*
)

sin



δ



sin

(
2

C
)
]
,
(
1
)
where. I
0
is the intensity of the incoming beam, E
x
and E
y
are the projections of the incident electric field vector parallel and perpendicular, respectively, to the transmission axis of the analyzer, &dgr; is the phase retardation of the compensator, C is the azimuth (rotational) angle of the fast (reference) axis of the compensator also relative to the transmission axis of the analyzer. In the case of a continuously rotating compesator, C=&ohgr;t, where &ohgr; is the angular rotational frequency of the compensator.
As can be seen by Eqn. (1), a rotating compensator will generate a signal having a dc component, a 2&ohgr; (two omega) component and a 4&ohgr; (four omega) component with respect to the rotation rate of the compensator. While usable information is generated from both the two omega and four omega signals, it is often felt that the two omega signal is the most significant for analysis. The two omega component is maximized when the phase retardation of the compensator is 90° (i.e. sin&dgr;=1), and disappears at phase retardations of 0° and 180° (sin&dgr;=0). Since the phase retardation of the compensator is a function of wavelength, this system lends itself to single wavelength operation only, or to a range where the wavelength changes only by a relatively small amount from the center wavelength of the compensator. As the wavelength deviates from the center wavelength such that the amount of phase retardation induced by the compensator deviates from 90 degrees, the relative intensity of the two omega signal is reduced. Therefore, multiple wavelength operation of rotating compensator ellipsometers has traditionally been limited to relatively narrow wavelength ranges (less than a factor of two in wavelength) corresponding to substantially 90 degree phase retardations induced by the compensator.
There is a need for a rotating compensator ellipsometer system that simultaneously obtains data over a wide range of wavelengths, which necessarily corresponds to phase retardations that vary significantly from the optimal 90°.
One prior-art solution to expand the effective wavelength range of retardation-based ellipsometers is to replace the rotating compensator with a photoelastic modulator. The phase retardation of the photoelastic modulator is a function of both the wavelength and the drive voltage to the modulator. During ellipsometric measurements, a series of wavelengths within the wavelength range of interest are sequentially scanned while the drive voltage to the photoelastic modulator is simultaneously changed. The change in drive voltage tracks the change in wavelength such that the phase retardation of the modulator is maintained at approximately 90°

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

Broadband spectroscopic rotating compensator ellipsometer does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with Broadband spectroscopic rotating compensator ellipsometer, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Broadband spectroscopic rotating compensator ellipsometer will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-2592849

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.