Time-of-flight ion mass spectrograph

Radiant energy – Ionic separation or analysis – Ion beam pulsing means with detector synchronizing means

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

C250S289000, C250S290000

Reexamination Certificate

active

06521887

ABSTRACT:

FIELD OF THE INVENTION
The present invention relates generally to mass spectrometry and, more particularly, to an apparatus that utilizes the detected position and detected time-of-flight of an electrostatically rastered ion beam to determine the speed and energy-per-charge ratio of individual ions from which the mass-per-charge ratio thereof can be derived. This invention was made with government support under Contract No. W-7405-ENG-36 awarded by the U.S. Department of Energy to The Regents of the University of California. The government has certain rights in the invention.
BACKGROUND OF THE INVENTION
Mass spectrometers are used to measure and analyze the chemical composition of substances. In general, they comprise an ion source, where neutral atoms or molecules from a solid, liquid, or gaseous sample are ionized, a mass analyzer where the atoms or molecules are separated according to their respective masses or their mass-per-charge ratios, and a detector. Various types of mass spectrometers exist, such as, for example, magnetic field spectrometers, quadrupole mass spectrometers, and time-of-flight mass spectrometers.
Magnetic sector mass spectrometers use a magnetic field or combined magnetic and electrostatic fields to measure the ion mass-per-charge. In one type of magnetic sector geometry (see, e.g., A. O. Nier, Rev. Sci. Instr. 18, 398 (1947) and L. Holmlid, Intl. J. Mass Spectrom. and Ion Phys. 17, (1975)), only one mass-per-charge species is detected at any time, so the magnetic field strength, and, if present, the electric field strength are varied to obtain a mass spectrum comprising of multiple mass-per-charge species. A major limitation of this type of sector mass spectrometer is the time that is required to scan the entire mass range one mass at a time.
In another type of magnetic sector mass spectrometer, ions are analyzed and focused onto a position-sensitive detector, resulting in the identification of ion mass-per-charge by spatial dispersion (see, e.g., J. Mattauch and R. Herzog, Zeitschrift fur Physik 89, 786 (1934)). While multiple mass-per-charge species can be detected simultaneously, the spatial resolution of the detector typically limits this type of spectrometer to a narrow mass range.
Quadrupole mass spectrometers utilize a mass filter having dynamic electric fields between four parallel electrodes (see, e.g.,
Quadrupole Mass Spectrometry and its Applications
, ed. Peter H. Dawson (American Institute of Physics, New York, 1995)). These fields are tailored to allow ions having a single mass-per-charge to pass through the filter at a time. One major limitation of quadrupole mass spectrometers is the time required to scan the entire mass range one mass at a time.
Time-of-flight mass spectrometry (TOFMS) can simultaneously detect ions over a wide mass range (see, e.g., M. Guilhaus, J. Mass Spectrom. 30, 1519 (1995)). Mass spectra are derived by measuring the times for ions to traverse a known distance. Generally, an ion's mass is derived in TOFMS by measurement or knowledge of an ion's energy E, measurement of the time t
1
that an ion passes a fixed point in space P
1
, and measurement of the later time t
2
that the ion passes a second point P
2
in space is located a distance d from P
1
. Using an ion beam having known energy-per-charge E/q, the time-of-flight (TOF) of the ion is equal to t
2
−t
1
, and the ion speed is v=d/(t
2
−t
1
). Since E=0.5 mv
2
, the ion mass-per-charge m/q is represented by the following equation:
m
q
=
2

E

(
t
2
-
t
1
)
2
qd
2
.
(1)
Implementation of a typical time-of-flight mass spectrometer known in the art is shown in
FIG. 1
hereof. A monoenergetic ion bunch,
10
, is introduced into a drift region,
12
, at time t
1
, and the time for ions from the front of the bunch to the back of the bunch to pass through the entrance aperture,
14
, is &Dgr;t
1
. The entire path the ions traverse is evacuated (not shown in FIG.
1
). Several types of gating exist to introduce an ion bunch with a small &Dgr;t
1
. For example, a pulsed, charged grid may be placed in front of aperture
14
to gate the ions entering drift region
12
, a pulsed laser can be utilized to generate bunches of ions, or an ion beam can be periodically deflected onto the entrance aperture using an electric field generated between two conducting plates,
16
, as shown in FIG.
1
. The ions traverse drift region
12
and disperse in space according to their speed and, therefore, mass, and impinge upon detector,
18
: the fastest, lightest ions are detected first, and the slower, heavier ions strike at a later time according to Equ. 1.
Conventional time-of-flight mass spectrometry has several major limitations.
First, the time during which the ion bunch transits the entrance aperture introduces an error, &Dgr;t
1
, in the mass-per-charge measurement; therefore, &Dgr;t
1
must be made short compared to the time-of-flight of the ions across the drift tube. Additionally, the lightest ions from a new bunch of ions can be admitted into the drift tube generally only after the heaviest ions from the preceding ion bunch are detected. While sophisticated techniques have been developed to overcome the limitation of overlapped spectra (see, e.g., U.S. Pat. No. 5,396,065 for “Sequencing Ion Packets For Ion Time-Of-Flight Mass Spectrometry” which issued to C. A. Myerholtz, et al. on Mar. 07, 1995, J. R. D. Copley, Nucl. Instr. and Meth. in Phys. Res. A291, 519 (1990), and G. Wilhelmi et al., Nucl. Instr. and Meth. in Phys. Res. 81, 36 (1970)), conventional time-of-flight mass spectrometers are inefficient since the duty cycle, which can be defined as the fraction of time that ions can enter the drift tube for analysis, is generally much less than unity. Second, it may be observed from Equ. 1 that the mass-per-charge ratio measurement is dependent on the distance d that the ion has traveled over the drift time t
2
−t
1
. Therefore, any errors in this time difference will produce corresponding errors in the derived mass-per-charge ratio. These errors are typically minimized by employing a long drift tube and a detector with a small detection area so that d is accurately known.
More recently, a non-time-of-flight technique has been used to determine the mass and velocity of ions. In U.S. Pat. No. 5,726,448 for “Rotating Field Mass and Velocity Analyzer” which issued to Steven Joel Smith and Ara Chutjian on Mar. 10, 1998, a rotating field mass and velocity analyzer having a cell with four walls, time dependent RF potentials applied to each of the walls and a detector is described. The time-dependent RF potentials create an RF field in the cell which effectively rotates within the cell. An ion beam accelerated into the cell is dispersed by the RF field according to the mass-to-charge ratio and velocity distribution of the ion beam. The ions of the beam either collide with the ion detector or deflect away therefrom depending on the mass-to-charge ratio, the RF amplitude, and the RF frequency. The detector counts the incident ions to determine the mass-to-charge ratio and the velocity distribution in the ion beam. Thus, ions that traverse a dynamic, (time-varying) electric field follow trajectories that are dependent on mass. Since the detector is located close to the cell (no time-of-flight tube), the resolution of the apparatus is limited by the time-of-flight of the ions.
Accordingly, it is an object of the present invention to provide an apparatus having a duty cycle of approximately unity for quantitatively measuring the speed, energy-per-charge, and mass-per-charge of ions.
Another object of the invention is to provide an apparatus for quantitatively measuring the speed, energy-per-charge, and mass-per-charge of ions where the deflected trajectory of an individual ion is independent of its mass.
Additional objects, advantages and novel features of the invention will be set forth in part in the description which follows, and in part will become apparent to those skilled in the art upon examination of the following or may b

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

Time-of-flight ion mass spectrograph does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with Time-of-flight ion mass spectrograph, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Time-of-flight ion mass spectrograph will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-3143612

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.