Method of performing a lasik procedure and tonometer system...

Surgery – Instruments – Light application

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

C606S005000, C128S898000, C600S398000

Reexamination Certificate

active

06730073

ABSTRACT:

BACKGROUND OF THE INVENTION
The present invention relates to an improved laser-assisted intrastromal in-situ keratomileusis (“LASIK”) procedure. More particularly, it relates to a LASIK procedure in which intraocular pressure is measured prior to cutting a flap in the cornea, along with a tonometer system capable of providing elevated intraocular pressure readings.
LASIK procedures have recently become highly popular for correcting various vision problems such as myopia, astigmatism, and hyperopia. In general terms, an automated microkeratome is employed to cut a large, superficial flap in the cornea. The flap is typically 160-180 microns in depth so as to include the first two corneal layers (i.e., the epithelium and Bowmans layer), while leaving the third layer or stroma in tact. The arc length of the so-defined flap is typically on the order of 300°, such that the flap can be folded back to expose the stroma. An excimer laser is then used to ablate or reshape the corneal stroma to effectuate the desired vision correction or enhancement. To this end, the excimer laser is typically centered on the prepupillary area. Following ablation, the flap is replaced or folded back. After a few minutes, the flap remains in place without requiring sutures, much like a suction cup.
As with other popular surgical procedures, numerous efforts have been made to improve the microkeratome and excimer laser instruments. One distinct aspect of the LASIK procedure, however, remains relatively rudimentary. In particular, to properly form the desired flap, the eyeball is subjected to a vacuum prior to cutting with the microkeratome. To this end, the microkeratome device includes a suction ring placed about the eye for applying the vacuum, resulting in a greatly elevated intraocular pressure as well as a slight distension of the eyeball from the socket. The elevated intraocular pressure renders the eyeball rigid, thereby promoting uniform, properly-sized incision by the microkeratome. Conversely, inadequate intraocular pressure can undesirably result in a thin, irregular flap.
As a point of reference, normal intraocular pressure is approximately 15-20 mm Hg. During a LASIK procedure, however, the intraocular pressure is preferably raised to a level in excess of 80 mm Hg to ensure proper eyeball rigidity. The optimal intraocular pressure for uniform microkeratome incision is currently unknown, due to the fact that available intraocular pressure sensing devices (or tonometers) have an upper limit at or below 80 mm Hg, as described in greater detail below. Thus, the currently accepted technique of evaluating intraocular pressure prior to microkeratome flap formation is known as a “finger” or “digit” method, whereby the surgeon touches the patient's eye with his/her finger and speculates as to the intraocular pressure. Obviously, evaluating intraocular pressure with the finger is highly subjective as the surgeon is required to guess as to whether the eye “feels” sufficiently rigid.
As indicated above, various tonometry instruments do exist for measuring intraocular pressure. Generally, there are two types of tonometers, including applanation and indentation. Either approach requires numbing of the eye and entails applying a force, via a probe placed against the cornea, that produces a distortion of the eyeball globe. The pressure required to applanate or indent the cornea is indicative of the intraocular pressure.
The most popular form of indentation tonometer is known as a Schiotz tonometer. With the Schiotz tonometer, a plunger produces a corneal indentation, the depth and volume of which are dependent upon the intraocular pressure. Other indentation tonometers, such as a Wolfe tonometer, rely on similar principals but with different internal or probe components, such as a calibrated spring.
An applanation tonometer measures the force required to flatten a certain area of the cornea. Examples of applanation tonometers include Goldmann tonometer, Draeger applanation tonometer, Mackay-Marg tonometer, and pneumatonometer, to name but a few. The Goldmann tonometer was the first example of a variable force applanation tonometer, and is still considered to be highly reliable. Other devices, such as the Mackay-Marg tonometer, measure intraocular pressure by analyzing deflections in membranes or crystals when the device is pressed against the eye. A pneumatonometer probe contains a gas and employs a pressure sensor or transducer. The pneumatonometer determines intraocular pressure by bringing a small burst of air toward the cornea. A backpressure is sensed, and is proportional to the intraocular pressure.
A third tonometry technique is generally referred to as “non-contact” and uses a stream of air (or “puff”) to flatten a portion of the cornea, while a light source is reflected off of that flattened portion. The pressure or force required to cause the sensed displacement is indicative of the intraocular pressure. While non-contact tonometers do not require anesthetizing the eye prior to use, they are generally regarded as being less accurate than applanation or indentation tonometers.
Each of the above-described tonometer systems were developed to detect the onset of glaucoma. Glaucoma is typically characterized by an increased intraocular pressure, above the normal 20 mm Hg. In this regard, a patient exhibiting a sensed intraocular pressure in excess of 30 mm Hg is typically deemed to be at risk for glaucoma; a 40-50 mm Hg intraocular pressure is highly abnormal, resulting in immediate emergency treatment. With these general parameters in mind, then, virtually all available tonometer devices have a maximum reading of approximately 50 mm Hg. In other words, for glaucoma testing, there is no need to provide for a larger range. Further, by limiting the range to approximately 50 mm Hg, the overall accuracy of the device can be increased.
At least one currently available pneumatonometer is able to accurately measure pressure up to 80 mm Hg. In particular, the Model 30 Classic™ pneumatonometer, available from Medtronic-Solan (formerly Mentor Ophthalmics) of Jacksonville, Fla., provides an intraocular pressure measurement in the range of 5-80 mm Hg. The Model 30 Classic™ device provides this enhanced intraocular pressure range to perform not only tonometer functions, but also tonography (i.e., evaluation of aqueous outflow of the eye). The accuracy of the Model 30 Classic™ is +/−2 mm Hg, and therefore is highly satisfactory for tonometery applications. It is believed that one other available tonometer has an upper limit of 90 mm Hg, but is limited to a simple “go or no-go” indication for intraocular pressure in excess of 50 mm Hg. Beyond these two examples, no device exists for measuring intraocular pressures in excess of 50 mm Hg.
The microkeratome and excimer laser instruments used for LASIK procedures continue to improve. However, the “finger touch” technique of evaluating intraocular pressure prior to microkeratome incision presents a distinct opportunity for complications. Therefore, a need exists for an improved LASIK procedure that promotes optimal flap formation and a tonometry system for effectuating this procedure.
SUMMARY OF THE INVENTION
One aspect of the present invention relates to a method of performing a LASIK procedure. The method includes providing LASIK instrumentation, preferably including an automated microkeratome instrument and an excimer laser. The microkeratome device includes both a microkeratome as well as a suction ring. A high-pressure range tonometer instrument is also provided. The tonometer includes a probe for sensing a parameter indicative of intraocular pressure, and is capable of providing intraocular pressure measurements in excess of 80 mm Hg; more preferably in excess of 100 mm Hg; most preferably up to or in excess of 120 mm Hg.
With these components in hand, the microkeratome instrument is positioned relative to the patient, including applying the suction ring over the patient's eye. A vacuum pressure is then applied to the eye via the suction ring. Th

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

Method of performing a lasik procedure and tonometer system... does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with Method of performing a lasik procedure and tonometer system..., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Method of performing a lasik procedure and tonometer system... will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-3232622

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.