Machine and process for the calibration and proving of flow...

Measuring and testing – Instrument proving or calibrating – Volume of flow – speed of flow – volume rate of flow – or mass...

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

C073S001190

Reexamination Certificate

active

06629447

ABSTRACT:

CROSS REFERENCE TO RELATED APPLICATIONS
Not Applicable
STATEMENT REGARDING FEDERALLY SPONSORED RESEARCH OR DEVELOPMENT
Not Applicable
BACKGROUND OF THE INVENTION
This invention relates generally to flow meter calibration and proving devices, and more specifically to a magnetically coupled, positive displacement, small volume, multi-axial, isokinetic, device that can be used as a flow calibrator or prover.
Flow calibrators or provers are used by metrology labs, flow sensor manufacturers, and field calibration operators. For users calibrating flow sensors requiring high accuracy, variable flow rates, extended flow turndowns, various calibrating fluids, and in some cases, fluctuating process conditions, the industry standard, positive displacement type calibrators/provers traditionally employed are: (1) The “sphere” calibrator/prover, often referred to as “ball” or “pipe” calibrator/provers, (2) The oscillating piston calibrator/prover, and (3) The “small volume” calibrator/prover (SVP). For custody transfer installations, these devices must provide an uncertainty of 1 part out of 10,000, as mandated in API Chapter 4, Manual of Petroleum Measurement Standards.
1) The “sphere” calibrator/prover typically uses over-inflated, spherical, elastomeric displacers, and requires large volumes of displaced calibration fluid per operation, due to the inherent repeatability inaccuracy of the mechanically activated, volume detector switches employed therein. They may be operated in a uni-, or bi-directional mode. They are typically constructed of commercial grade, carbon steel, pipe and fittings.
2) The oscillating piston type of calibrator/prover typically employs a solid metal, free floating piston, with “cup” seals, that oscillate between detector switches and is similar to the “sphere” type, in construction, and large volume displacement required.
3) The SVP type of calibrator/prover typically utilizes an internal metal poppet valve mounted in a precision machined steel piston, with special seals, oscillating in a plated, honed, precision machined, cylindrical flow tube. The piston is retracted upstream between calibration cycles by means of a pneumatic, hydraulic, or mechanical coupling of the piston shaft. The displaced volume of the SVP is a fraction of that for the “sphere” or oscillating piston types due to the use of precision volume detector switches, dual megahertz chronometers, and pulse interpolation of the test sensor's output. The requirements for the SVP, are defined in the API, Manual of Petroleum Measurement Standards, Chapters 4.3, and 4.6.
The moderate to high cost of “sphere”, oscillating piston, and SVP type calibrator/provers, typically restricts the use of these devices to those users, and service contractors, involved in custody transfer of hydrocarbon liquids, manufacturers of large quantities of flow sensors, and commercial metrology facilities. Start/stop operation of the displacer is common to all of these devices. Other inherent disadvantages are:
1) “Sphere” and oscillating piston types require a large amount of dedicated, cubic operating space, and high volume storage tanks for calibrating fluids, in the calibration lab. Cost of ownership is considerable, including maintenance and periodical NIST traceable recertification. The use of various calibrating fluids is restricted by the materials of construction of the device. Due to large dispensed volumes required, multi-pass calibrations, which require start/stop of the displacer, and those performed at low flow rates can be extraordinarily lengthy.
2) SVP types achieve the industry mandated accuracy, with the use of precision electronics, pulse interpolation, and special proprietary seals, operating in a precision honed cylinder. The cylinder is typically centrifugally cast or forged. Care must be exercised to prevent these surfaces from corrosion and minute defects. Energy and time must be expended to retract the oscillating piston upstream, against the fluid flow, between calibration cycles. Another inherent disadvantage of the SVP, is the inability to accurately calibrate flow sensors with minimal or irregular pulse counts per unit of displaced volume, due to the small amount of displaced fluid per calibration cycle. Also, mass flow meters such as the coriolis type not only require larger amounts of displaced fluid per cycle, but are affected by the flow perturbations induced when the poppet valve closes to launch the piston downstream. When the SVP is equipped with the typical up and downstream volume switches, the dispensed volume represented by the spacing of these two switches must also be displaced through small flow sensors that could be calibrated with much less volume displaced. Cost of ownership is typically less than that for “sphere” types, but still considerable.
BRIEF SUMMARY OF THE INVENTION
A primary object of the present invention is to provide a magnetically coupled, small volume, positive displacement, fully automatic, flow sensor calibrator/prover.
Another object of the invention is to provide a high accuracy, economical calibrator/prover, with increased operating speed, that requires minimally dispensed calibration fluid, and a small operating space, without precision machined, proprietary components.
Yet another object of the invention is to perform single or multi-pass, calibration cycles by isokinetically dispensing increments of the composite volume displaced during a single, unidirectional transit of the displacer.
In accordance with a preferred embodiment of the present invention, ‘The invention includes a single, or multiple serially connected, flow tube(s), the number of which determines the number of calibration cycles performed, with one unidirectional passage of a free floating displacer, from initiation to acquisition in an Inline Automatic Launch/Retrieve Station (LRS), means a discrete, pre-calibrated, displaced volume of calibration fluid for each flow tube, equating to spacing between Volume Detector Sensors (VDS), is isokinetically dispensed by the displacer through test meters(s), while simultaneously, the electronic outputs from the VDS, the test meter(s), and optional auxiliary instruments, are interpolated, and processed in the P.C. or PLC based operator Monitoring & Control Station(MCS), means a cylindrical, convoluted, magnetically coupled displacer, with anti-compression device(s), and imbedded magnet(s), which emits an omni-directional magnetic field, is sequentially coupled to, and decoupled from, a coaxial, or parallel, oscillating, Volume Detector Actuator (VDA), means the calibration fluid flow propels the displacer, in one direction, through a flow tube, then a return bend, and subsequently through the next flow tube, in the opposite direction, reacquiring the outer ring of the VDA, to initiate the calibration cycle for that flow tube, means each outer ring of the VDA, is located concentrically,or parallel, to a flow tube, and is connected to a common, circular center ring, simultaneously oscillating with the outer ring(s), and traversing coaxial to a centrally mounted Volume Detector Rod (VDR), said VDA center ring sequentially activates the VDS's, by means of a VDS Initiator, a projection integral to the VDA, and maintains dynamic concentricity by virtue of integral centering devices located in the VDA center ring.
The inclusion of a convoluted, cylindrical displacer in the apparatus, in lieu of the spherical type previously employed, allows the displacer to traverse pipeline tees, and standard, long radius elbows. The omni-directional imbedded magnet (s), enables the displacer to magnetically couple to, and decouple from, a coaxial, or parallel, oscillating, VDA. The convolutions of the displacer provide multiple, dynamic, seal surfaces, separated by liquid filled annular spaces, and eliminates inner poppet seals of the “small volume” piston type prover(SVP). This combination of elements eliminates the start/stop operation of previous designs, and provides multiple, sequential, calibration cycles, for multiple sensors, with a single, unidirection

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

Machine and process for the calibration and proving of flow... does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with Machine and process for the calibration and proving of flow..., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Machine and process for the calibration and proving of flow... will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-3139661

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.