Laser eye surgery system using wavefront sensor analysis to...

Surgery – Instruments – Light application

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

C606S004000, C356S002000, C351S212000, C128S898000

Reexamination Certificate

active

06394999

ABSTRACT:

BACKGROUND OF THE INVENTION
1. Field of the Invention
This invention relates broadly to eye surgery. More particularly, this invention relates to refractive laser systems for eye surgery.
2. State of the Art
The laser refractive surgery (or laser keratectomy) field has exploded over the past few years with many new lasers and algorithms to correct human vision. Systems are now using laser wavelengths from the ultraviolet (excimer) to the infrared to change the shape of the cornea in a calculated pattern which makes it possible for the eye to focus properly. For example, in the treatment of myopia, the excimer laser is used to remove or ablate tissue from the cornea in order to flatten its shape. Infrared (IR) energy is also used by some companies to treat myopia by reshaping the corneal tissue by a “thermal” method as opposed to ablation with the excimer wavelength. The correction of hyperopia is produced by steepening the cornea by removing tissue at the outer edges of the cornea (excimer) or by reshaping the cornea at the outer edges (IR energy). The correction of astigmatism, both myopic and hyperopic, requires the laser to remove (as done by the excimer laser) or reshape (as done by the IR laser) tissue in a more complex pattern. Laser corneal reshaping procedures are effective for correcting impaired visual acuity, but many researchers now believe this effectiveness needs to be improved. The proper model and delivery of the laser energy for removing or altering the tissue has been a major discussion recently as postoperative studies are indicating that current procedures may actually be inducing aberrations in the eye optical system.
Initial systems approved by the FDA implement the refractive corrections by a broadbeam approach; i.e., by delivering beam-shaped laser energy based on thin lens theory and paraxial optics applied to a single spherical surface. The beam is shaped by a motorized iris (myopia and hyperopia) and motorized slit (astigmatism) based on profiles derived through Munnerlyn's derivation (C. R. Munnerlyn, S. J. Koons, and J. Marshall, “Photorefractive keratectomy: a technique for laser refractive surgery”,
J. Cataract Refract. Surg.
14, 46-52 (1988)). Systems using this approach are currently marketed by VISX and Summit. More than one million eyes have been treated in this manner in the United States. However, this approach is limited, as it symmetrically treats a broad area of the cornea all at one time. Eye topography maps and, more recently, wavefront analysis reveal that the cornea is a very complex structure with many minute variations across its surface. The broadbeam laser approach cannot correct these minute variations.
A more recent approach to laser keratectomy uses a scanning laser spot system in which a small laser spot (typically 0.5-mm to 1.0-mm in diameter) is scanned across the cornea in a predetermined pattern to achieve refractive corrections. These systems differ in that they are more flexible than the broadbeam approach. With the control of a small spot, different areas of the cornea can be shaped independently of other areas. The scanning spot approach, therefore, allows for the laser beam to be moved in a specific pattern over the cornea to more particularly correct the shape of the cornea. However, there are several problematic issues with scanning spot systems.
First is the issue of treatment time. Scanning spot systems require longer refractive surgery times. The scanning spot is a slower approach since the small laser spot has to be moved over a wide surface (up to 10-mm for hyperopia). The scanning spot system typically delivers several hundred spots per treatment layer, and consequently treatment times are generally long. The broadbeam approach is much quicker as the entire cornea is treated with each laser pulse, or treatment layer.
Second is the issue of safety. The broadbeam laser is inherently safe from a treatment interruption standpoint because the cornea is treated symmetrically for each pulse; the iris represents a circle and the slit represents a rectangle so that every point on the cornea being treated is treated the same with each laser pulse. If the procedure is interrupted, there will always be some symmetrical spherical correction which can be continued more easily at a later time. However, the scanning spot, with its small spot size, cannot cover the entire corneal surface with one laser pulse. Thus, if an interruption occurs, there is no guarantee of a complete corneal etch for a layer at the point of interruption. Continuation at the point of interruption would be difficult.
Third is the issue of tracking. In the scanning spot system the eye needs to be tracked very carefully in order to deliver the spot to the correct point on the cornea as the eye moves. This is not as much of a problem in the broadbeam system as a broader area is treated with each pulse.
Fourth is the issue of surface roughness. The necessary overlap of laser spots tends to create roughness in the resulting etch. While it is necessary to overlap spots to provide complete coverage for a given ablation zone, regions of overlap will be ablated at twice the etch depth per pulse. The smoothness of the ablated volume is dependent on the spot overlap and to a lesser extent, the ratio of spot diameter and ablation zone diameter. This problem is not seen in the broadbeam approach.
Current FDA-approved refractive laser systems do not directly use eye-modeling systems, such as corneal topographers or wavefront sensors, to create an individual treatment profile for a patient's eye. Rather, a topographic map is used indirectly by the surgeon in optimizing the treatment plan (diopter correction and astigmatic axis). More recently, eye contour topography has been used to more accurately provide refractive correction. There are systems currently going through FDA trials that do use corneal topographic surface data to directly guide the laser treatment algorithm. This is accomplished using a scanning laser spot, as current broadbeam laser refractive surgery systems cannot provide the laser beam detail required to use the topographic map data. Using the scanning spot system, each eye is individually analyzed as to its contour before ablation is applied. The goal with these systems is to take into account the varying degrees of curvature and height variations across the corneal surface, as opposed to assuming a perfect spherical surface as is currently done in broadbeam systems. Once these curvatures and height variations are determined by eye topography, they can be integrated with the refraction correction derivation of Munnerlyn to create customized ablation patterns for each individual eye.
The corneal topography approach has a drawback in that only measurements of the cornea are used. However, the eye is a complex optical system of which the cornea is only one component. Thus, even corneal topography information when combined with the current FDA-approved refraction equation, is not capable of suggesting what correction must be made to the corneal shape in order to optimally correct the overall aberration of the eye's optical system.
There have been several recent approaches to the above problems. First, by expanding the mathematical equations for refraction correction to include higher order effects, coma (3rd order) and spherical (4th order) aberrations can be reduced. See C. E. Martinez, R. A. Applegate, H. C. Howland, S. D. Klyce, M. B. McDonald, and J. P. Medina, “Changes in corneal aberration structure after photorefractive keratectomy,”
Invest. Ophthalmol. Visual Sci. Suppl.
37, 933 (1996). Second, by improving schematic model eyes to include higher order aberrations, these new models can provide insight into how the various elements of the eye optical system correlate to affect visual performance. For example, there is a general consensus that the negative asphericity of the normal cornea contributes a negative aberration content. The negative aberration is compensated by a positive aberration contribution from the gradient i

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

Laser eye surgery system using wavefront sensor analysis to... does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with Laser eye surgery system using wavefront sensor analysis to..., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Laser eye surgery system using wavefront sensor analysis to... will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-2901362

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.