DEVICE AND SYSTEM FOR REMOTE FOR IN-CLINIC...

Surgery – Diagnostic testing – Detecting muscle electrical signal

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

C600S304000, C600S509000, C600S511000

Reexamination Certificate

active

06816744

ABSTRACT:

SPECIFICATION
This application claims the benefit of U.S. Provisional Application Ser. No. 60/294,391, filed May 29, 2001.
FIELD OF THE INVENTION
The present invention relates to measurements of electrical activity in a body. More particularly, the invention relates to measurement and analysis of the measurements to predict the condition of a portion of a body.
BACKGROUND OF THE INVENTION
Presently there is no objective manner with which to evaluate the contractility of the uterus. This is true either in non-pregnant patients where hypercontractility is associated with dysmenorrhea or in pregnant patients where the uterus is sometimes active prior to term. Normally the uterus is quiescent in non-pregnant women and during most of pregnancy. However, at the end of pregnancy, the myometrium undergoes a series of changes that lead to synchronous, rhythmic uterine contractions (labor). The diagnosis of labor is the most significant problem faced by obstetricians. In addition, pre-term labor, which occurs in about 10% of pregnant patients, is difficult to diagnose. Frequently, term or pre-term labor requires adjuvant therapy to either stimulate or inhibit contractility of the uterus.
Since there is some minor spontaneous uterine activity at all times during pregnancy, it is often not possible to distinguish between this physiological activity at term or preterm labor. The state of the cervix is commonly used as a predictor of labor. However, the dilatation of the cervix usually occurs relatively late, during actual labor. In addition, labor and changes in the cervix can occur independently. Alternatively the frequency of contractions is used to diagnose labor, sometimes recorded with a tocodynamometer. However, these methods give only crude subjective estimates of uterine contractility.
The uterus does not contract vigorously throughout most of pregnancy and this provides a tranquil environment for the growing fetus. At term, the uterus normally begins to contract forcefully in a phasic manner (labor) to expel the fetus. Contractions of the uterus are directly proportional to the underlying electrical activity of the muscle. The frequency, duration, and magnitude of a uterine contraction are directly proportional, respectively, to frequency of bursts of action potentials, and the propagation (also referred to as conduction) of action potentials over the uterus and the recruitment of muscle cells. A similar situation exists in heart muscle, although heart and uterine muscle are different with respect to structure and configuration of the action potentials. The action potentials are accompanied by the influx of calcium into the muscle cells to activate the contractile apparatus.
Thus, by recording uterine electrical activity one can assess the contractility of the myometrium. Similar technology is used to record cardiac electrical activity to determine the normal or abnormal function of the heart.
Many studies have previously recorded myometrial electrical activity using electromyography (EMG) where electrodes are placed directly on the uterus. These studies show that the myometrium generates little electrical activity prior to labor but activity increases tremendously during labor reflecting the mechanical events. Studies of interest are demonstrated in publications by Csapo, Chapter 43, “Force of Labor,”
Principles and Practice of Obstetrics and Perinatology
, Ed. L. Iffy and H. A. Kaminetzky, John Wiley and Sons Publishing 761-799, 1981; Garfield et al., “Control of Myometrial Contractility: Role and Regulation of Gap Junctions,” Oxford Rev. Reprod. Biol. 10:436-490, 1988; Wolfs and Van Leeuwen, “Electromyography observations on the human uterus during labor,” Acta Obstet. Gynecol. Scand. [Suppl.] 90:1-62, 1979; and more recently by Devedeux et al., “Uterine Electromyography: A Critical Review,” Am J. Obstet. Gynecol. 169:1636-1653, 1993. One may measure and use uterine EMG activity by direct contact with the uterus to predict normal and abnormal uterine contractions. However, it is not practical to place electrodes directly on the uterus. To do this under the present level of understanding one must surgically implant electrodes on the uterine surface or introduce a catheter electrode through the vaginal canal and puncture the fetal membranes.
It would be desirable to record uterine EMG activity from the abdominal, cervical or vaginal surface. Previous studies of electrical activity of the uterus recorded with electrodes placed on the abdominal surface failed to record bursts of action potentials from the uterus and generally showed no association of uterine electrical activity with contractility. Studies of interest are included in the above-noted publications by Wolfs and Van Leeuwea and by Devedeux et al. Wolfs and Van Leeuwea summarized all studies prior to 1979 and concluded, “it has never been clearly shown that the potential fluctuations obtained by means of electrodes attached to the abdominal wall, do indeed represent the electrical activity of the uterus.” (Page 7.) Similarly, Devedeux et al state that abdominal monitoring of uterine electrical activity “requires further investigation” (Page 1649).
Recently, studies have been done which establish that there is significant correlation between the potentials of the uterus as measured at the abdominal surface and directly at the uterus. These studies show that such electrical signals can be quantified by mathematical means, for example, with Fourier analysis or Wavelet analysis: Garfield, R E, et al, “Control and assessment of the uterus and cervix during pregnancy and labour, 1996”; Buhimschi C, Garfield R E. “Uterine activity during pregnancy and labor assessed by simultaneous recordings from the myometrium and abdominal surface in the rat,” Am. J. Obstet Gynecol 1998, 178:811-22; and Garfield R E, et al, “Instrumentation for the diagnosis of term and preterm labour,” J. Perinat Med 1998; 26; 413-436.
Part of the difficulty in interpretation of electrical activity recorded from the uterus lies in the fact many investigators, including Wolfs and Van Leeuwea and Devedeux et al., have failed to recognize that action potentials drive the uterus to contract. Action potentials are not responsible for contraction of some smooth muscle tissues such as airway muscle and some vascular muscles, and therefore many researchers confound the uterus with other smooth muscles. Thus, many of these studies have attempted to correlate electrical activity with mechanical contractions in order to show that electrical activity is or is not responsible for contractions. It is now clear (from publications by Marshall, “Regulation of Activity in Uterine Smooth Muscle,” Physiol. Rev. 42-212-227, 1962; Csapo, Chapter 43. “Force of Labor,”
Principles and Practice of Obstetrics and Perinatology
, Ed. by I. Iffy and H. A. Kaminetsky, John Wiley & Sons Publishing, 761-799, 1981; Garfield et al., “Control of Myometrial Contractility: Role and Regulation of Gap Junctions,” Oxford Rev. Reprod. Biol. 10:436-490, 1988; and Garfield, Chapter 3 “Role of cell-to-cell Coupling in Control of Myometrial Contractility and Labor,”
Control of Uterine Contractility
, Ed. R. E. Garfield and T. Tabb, CRC Press. 39-81. 1994) that action potentials activate the uterus to contract and that by measuring uterine electrical activity one can indirectly estimate contractility.
There has been much progress in monitoring adult ECG using an array of surface electrodes placed on the skin (
Interventional Electrophysiology
, 2
nd
Edition, Ed. by Singer. Lippincott Williams & Wilkins, April 2002). There has also been some success in monitoring maternal and fetal cardiac activity from the abdominal surface of pregnant patients using electrodes (Kanjilal, et. al., “Fetal ECG Extraction from Single-Channel Maternal ECG Using Singular Value Decomposition,” IEEE Trans Biomed Eng. January; 44(1): 51-9, 1997; Kwon, et. al., “Abdominal Fetal EKG Noise Removal,” Biomed Sci Instrum. 32: 87-92, 1996). However, the devices and methods of data acquisition and signal processing delin

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

DEVICE AND SYSTEM FOR REMOTE FOR IN-CLINIC... does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with DEVICE AND SYSTEM FOR REMOTE FOR IN-CLINIC..., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and DEVICE AND SYSTEM FOR REMOTE FOR IN-CLINIC... will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-3330177

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.