Defibrillation shock strength determination technology

Surgery: light – thermal – and electrical application – Light – thermal – and electrical application – Electrical therapeutic systems

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Reexamination Certificate

active

06675042

ABSTRACT:

STATEMENT REGARDING FEDERALLY SPONSORED RESEARCH OR DEVELOPMENT
Not applicable.
REFERENCE TO A MICROFICHE APPENDIX IF ANY
Not applicable.
BACKGROUND OF THE INVENTION
1. Field of the Invention
The present invention relates, generally, to implantable cardioverter defibrillators (ICDs) and defibrillation methods, and particularly to a method and apparatus for determining the optimal shock strength for defibrillation, and most particularly to determining the upper limit of vulnerability (ULV) based on changes with respect to time in the T-wave of the cardiac signal, preferably the maximum of the first derivative of the T-wave with respect to time measured preferably exclusively from implanted electrodes. Unless otherwise indicated, the term “derivative of the T-wave” refers to the first derivative of the T-wave with respect to time. The technology is useful for automating the process of selecting the first defibrillation shock strength for ICDs.
2. Background Information
Heart disease is a leading cause of death in the United States. The most common form of cardiac death is sudden, caused by cardiac rhythm disturbances (arrhythmias) in the form of a ventricular tachycardia or ventricular fibrillation.
Ventricular tachycardia is an organized arrhythmia originating in the ventricles. It results in cardiac contractions that are too fast or too weak to pump blood effectively. Ventricular fibrillation is a chaotic rhythm disturbance originating in the ventricles that causes uncoordinated cardiac contractions that are incapable of pumping any blood. In both ventricular tachycardia and ventricular fibrillation, the victim will most likely die of “sudden cardiac death” if the normal cardiac rhythm is not reestablished within a few minutes.
Implantable cardioverter defibrillators (ICDs) were developed to prevent sudden cardiac death in high risk patients. In general, an ICD system consists of implanted electrodes and a pulse generator that houses implanted electrical components. The ICD uses implanted electrodes to sense cardiac electrical signals, determine the cardiac rhythm from these sensed signals, and deliver an electrical shock to the heart if life-threatening ventricular tachycardia or ventricular fibrillation is present. This shock must be of sufficient strength to defibrillate the heart by simultaneously depolarizing all or nearly all heart tissue.
Shock strength is typically measured as shock energy in Joules (J). The defibrillating shock interrupts the abnormal electrical circuits of ventricular tachycardia or ventricular fibrillation, thereby permitting the patient's underlying normal rhythm to be reestablished. ICD pulse generators are implanted within the patient and connected to the heart through electrodes to provide continuous monitoring and immediate shocking when a life-threatening rhythm disturbance is detected. Because the devices must be small enough for convenient implantation, ICDs are limited in their ability to store electrical energy. In general, ventricular tachycardia can be terminated by weaker shocks than those required to terminate ventricular fibrillation. Thus ICDs must deliver a sufficiently strong shock to insure reliable defibrillation in response to each occurrence of ventricular fibrillation.
One method is to use the maximum shock strength of the ICD for each shock. However, this approach is an inefficient use of the ICD's limited stored electrical energy and will unnecessarily reduce the useful life of an ICD pulse generator.
It is well known in the art that the shock strength required to defibrillate a human heart effectively varies with the implanted lead configuration and placement as well as the individual heart's responsiveness to the shock. To maximize efficiency of an ICD system, the minimum shock strength necessary to defibrillate an individual patient's heart reliably must be determined.
However, it is also well known in the art that the relationship between an ICD's defibrillation shock strength and success or failure of defibrillation is represented by a probability-of-success curve rather than an all-or-none defibrillation threshold (DFT). Very weak, low strength (low energy) shocks never defibrillate. Very strong shocks, at energies greater than the maximum output of ICDs, always defibrillate. However, clinically relevant shock strengths for ICDs lie between these two extremes. In this intermediate range of shock strengths, a shock of a given strength may defibrillate successfully on one attempt and not on another attempt.
Determining a complete curve of the probability of success for every possible shock strength requires many fibrillation-defibrillation episodes. In clinical (human) studies and procedures, the number of fibrillation-defibrillation episodes should be limited because of their associated risks. Thus the goal of testing at the time of ICD implant cannot be to determine a complete probability of success curve. In general, the goal of testing at ICD implant is to provide an accurate estimate of the minimum shock strength that defibrillates with a high probability of success while using a minimum amount of testing. The shock energy that defibrillates with an X% probability of success is referred to as the defibrillation threshold
x
or DFT
X
. Thus a goal of clinical testing at ICD implantation is to estimate a shock strength in the range of the DFT
95
-DFT
99
. This is the optimal strength at which to program the first shock of an ICD. For research purposes, it may be preferable to estimate the DFT
50
.
The minimum measured shock strength that defibrillates during a given session of defibrillation testing is referred to, in general, by the term DFT, despite the fact that no true threshold for defibrillation exists. All methods for determining the DFT of an ICD system require inducing fibrillation a number of times and testing various shock strengths for defibrillation through the implanted defibrillation leads. In the commonly used step-down method defibrillation is attempted at a high shock strength that is likely to defibrillate the heart successfully. If this shock is unsuccessful, a stronger “rescue shock” is delivered to effect defibrillation. Regardless of the outcome of the defibrillation shock, there is a waiting period of about 5 minutes to permit the patient's heart to recover. If the defibrillation shock is successful, fibrillation is reinitiated and the defibrillation is attempted at a lower shock strength. This process is repeated with successively lower defibrillation shock energies until the shock does not defibrillate the heart. The minimum shock strength that defibrillates is the DFT. Depending on the initial shock strength, the DFT determined in this manner is usually between the DFT
30
and DFT
70
. The ICD is then programmed to a first-shock strength selected to be an estimate of the lowest value that can reliably achieve defibrillation by adding an empirically-determined safety margin to the DFT.
Other methods for determining the DFT require additional fibrillation-defibrillation episodes after a defibrillation shock has failed. In these methods, fibrillation is reinitiated after a failed defibrillation shock and defibrillation is attempted at successively higher shock strengths until a shock defibrillates the heart successfully. This change from a shock strength that does not defibrillate to one that does (or vice versa) is called a reversal of response. DFT methods may require a fixed number of reversals. If the size of the shock increments and decrements is the same, a multiple-reversal (up-down) method provides a good estimate of the DFT
50
. An alternative Bayesian method uses a predetermined number of unequal shock increment steps and decrement steps to estimate an arbitrary, specific point on the DFT probability of success curve.
One significant disadvantage of all DFT methods is the necessity to repeatedly fibrillate and then defibrillate the patient's heart to determine the DFT. For example, U.S. Pat. No. 5,531,770 describes a method of DFT testing that is de

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

Defibrillation shock strength determination technology does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with Defibrillation shock strength determination technology, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Defibrillation shock strength determination technology will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-3245697

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.