MS/MS scan methods for a quadrupole/time of flight tandem...

Radiant energy – Ionic separation or analysis – Ion beam pulsing means with detector synchronizing means

Reexamination Certificate

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Reexamination Certificate

active

06507019

ABSTRACT:

FIELD OF THE INVENTION
This invention relates to mass spectrometry including multiple mass analysis (MS/MS) steps and final analysis in a time of flight (TOF) device or in general any orthogonal mass spectrometry system. This invention is more particularly concerned with such a technique carried out in a hybrid tandem quadrupole-TOF (QqTOF) spectrometer and is concerned with improving the duty cycle of such an instrument for parent or precursor ion scanning and like operations, or more generally to improving the duty cycle over a wide mass range for any type of scan.
BACKGROUND OF THE INVENTION
Tandem mass spectrometry is widely used for trace analysis and for the determination of the structures of ions. In tandem mass spectrometry a first mass analyzer selects ions of one particular mass to charge ratio (or range of mass to charge ratios) from ions supplied by an ion source, the ions are fragmented and a second mass analyzer records the mass spectrum of the fragment ions. In a triple quadrupole mass spectrometer system, this effects MS/MS. Ions produced in an atmospheric pressure source, pass through a region of dry nitrogen and then pass through a small orifice into a region at a pressure of several torr. The ions then pass through, a quadrupole ion guide, operated a pressure of about 7×10 3 torr into a first quadrupole mass analyzer, operates at a pressure of about 2×10−5 torr. Precursor ions mass selected in the first quadrupole mass analyzer are injected into a collision cell filled with an inert gas, such as argon, of a pressure of 10
−4
to 10
−2
torr. The collision cell contains a second quadrupole (or multipole) ion guide, to confine ions to the axis. Ions gain internal energy through collisions with gas and then fragment. The fragment ions and any undissociated precursor ions then pass into a third quadrupole, which forms a second mass analyzer, and then to a detector, where the mass spectrum is recorded.
Triple quadrupole systems are widely used for tandem mass spectrometry. One limitation is that recording a fragment mass spectrum can be time consuming because the second mass analyzer must step through many masses to record a complete spectrum. As in any scanning mass analyzer, all other ions (outside of ‘transmission window’) are lost for analysis, thus reducing the duty cycle to values of around 0.1% or less. To overcome these limitations, QqTOF systems have been developed (as described for example in: Morris, H. R.; Pacton, T; Dell, A.; Langhorne, J.; Berg. M.; Bordoli, R. S.; Hoyes, J.; Bateman, R. H.;
Rapid Commun. Mass Spectrometry,
1996, 10, 889-896; and Shevohenko, A.; Chernushevich, I.; Ens, W.; Standing, K. G.; Thomson, B.; Wilm, M.; Mann, M.,
Rapid Commun. Mass Spectrometry,
1997, 11, 1015-1024). This system is similar to the triple quadrupole system but the second mass analyzer is replaced by a time-of-flight mass analyzer, TOF. The advantage of the TOF is that it can record 10
4
or more complete mass spectra in one second without scanning. Thus for applications where a complete mass spetrum of fragment ions is desired the duty cycle is greatly improved with a TOF mass analyzer and spectra can be acquired more quickly. Alternatively for a given measurement time, spectra can be acquired on a smaller amount of sample.
A further known technique is the coupling of electrospray ionization (ESI) to time-of-flight mass spectrometers (TOFMS), and this is an attractive technique for mass spectrometry. ESI is a soft ionization technique capable of forming ions from a broad range of biomolecules, while TOFMS has the well known advantages of rapid mass scanning, high sensitivity, and a theoretically limitless mass range. However, ESI and TOFMS are, in one way, incompatible as a source/analyzer pair: ESI creates a continuous stream of ions and TOFMS requires pulsed operation. Thus in the simplest coupling of ESI to TOFMS there is a very poor duty cycle, with less than 1% of the ions formed being detected (to obtain reasonable mass resolution) and early work in this field was predominantly concerned with increasing the duty cycle.
Within the past two years, literature on ESI-TOFMS has begun to focus on tandem mass spectrometry (MS/MS) with hybrid instruments. The fragmentation of ions in these systems is achieved via traditional methods for collision induced dissociation (CID), Tandem-in-space systems termed quadrupole-TOF's (QqTOF of QTOF), as noted above, are analogous to triple quadrupole mass spectrometers—the precursor ion is selected in a quadrupole mass fitter, dissociated in a radiofrequency- (RF-) only multipole collision cell, and the resultant fragments are analyzed in a TOFMS. Tandem-in-time systems use a 3-D Ion trap mass spectrometer (ITMS) for selecting and fragmenting the precursor ion, but pulse the fragment ions out of the trap and into a TOFMS for mass analysis.
Tandem mass spectrometers (in particular, triple quadrapoles and QqTOFs) are often used to perform a technique known as a parent ion scan (or precursor ion scan). In this techniques the first mass resolving quadrupole is scanned in order to sequentially transmit precursor ions over a selected mass range. The second mass spectrometer is used to selectively transmit only one specific fragment or product ion from the collision cell. The mass spectrum thus produce by scanning, the first mass spectrometer shows only those ions from the ion source which fragment to produce the specific product ion. Thus from a complex mixture of ionized species, a simple mass spectrum allowing only those components which produce the known fragment ion is produced. This method is often used in order to identify precursor ions as candidates for fill MS/MS. For example, if the sample contains a mixture of many different species, and the only compounds of interest are those which have a structure known to always generate a fragment of m/z 86, then a precursor ion scan may be performed in order to identify which precursor ions form m/z 86. A full MS/MS spectrum may then be performed on those few precursor ions, instead of on every peak in the Q1 mass spectrum. In this way, a significant amount of time can be saved in analyzing the sample.
In triple quadrapoles, precursor ion scans have proved to be the right tool to search for ions of certain classes of compounds, e.g. peptides
1
, glycopeptides
2
or phosphopeptides
3
(as detailed, for example in the following references for these three classes of compounds,
1
M Wilm, G. Neubauer and M. Mann,
Anal. Chem.,
1996 88, pp. 527-633;
2
S. A. Carr, M. J. Huddleston and M. F. Bean,
Protein Science,
1993, 2, pp.183-198;
3
S. A. Carr, M. J. Huddleston and R. S. Annan,
Anal. Biochem.,
1996, 239, pp 180-192). However, a current limitation of the Qq-TOFs is their lower sensitivity in this particular mode of operation, compared to triple quadrupoles. The last mass analyzer (TOF or Q3) does not need to scan in this mode, and the Qq-TOF does not benefit from simultaneous ion detection in TOF. On the other hand, more ions are lost in a TOF compared to a third quadrupole: at the entrance, on grids, and mostly due to duty cycle.
The problem here is that usually the fragment ions cover a large m/z range, and the TOF instrument has to capture all that m/z range if consecutive spectra are not to overlap. If one is interested in just a particular mass, then this can lead to a low duty cycle.
There are two main factors governing the duty cycle of an orthogonal acceleration TOF instrument when operated in the conventional (continuous beam) mode. Generally, you have to wait for the heaviest ions to reach the detector before the next pulse of ions can be introduced. Since the width of the entrance window is only a traction of the transverse distance between the ion storage region and the detector, even the heaviest ions will overfill this region before the next pulse of ions can be released. The loss due to this effect is simply equal to the ratio of the length of the entrance window to the distance between the storage region and the detector.

LandOfFree

Say what you really think

Search LandOfFree.com for the USA inventors and patents. Rate them and share your experience with other people.

Rating

MS/MS scan methods for a quadrupole/time of flight tandem... does not yet have a rating. At this time, there are no reviews or comments for this patent.

If you have personal experience with MS/MS scan methods for a quadrupole/time of flight tandem..., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and MS/MS scan methods for a quadrupole/time of flight tandem... will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFUS-PAI-O-3007984

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.